7. Select the start field.
Result
A ventilation mode for a new patient has been set. If the curve
display option is activated, the device shows a curve view.
4.7.4 Selecting an emergency mode from a
ventilation mode
Requirement
•
•
1. Select the field for the emergency mode using the right-hand
2. Select emergency mode:
The device is switched on.
A ventilation mode is set (exception: CPR, Inhalation, CO
monitoring)
navigation knob.
•
Emergency Infant
•
Emergency Child
•
Emergency Adult
Depending on the preset in the operator menu, the device
switches to one of the following modes with the ventilation
parameters preset for the patient group
settings for emergency modes and ventilation modes",
page 234)
and shows a pressure gauge view:
4 Preparation and operation
(see "14.1.9 Factory
MEDUMAT Standard
2
2
EN
73