7 Description of the modes

7.2 Ventilation parameters

138
EN
MEDUMAT Standard
Ventilation
parameters
Vt
Freq.
pMax
PEEP
Air Mix
93% oxygen
100% oxygen
Δ pASB (only with
flow measurement +
ASB option)
InTr level (only if the
flow
measurement + ASB
option is activated)
ExTr level (only if the
flow measurement +
ASB option is
activated)
InTr (only if the flow
measurement + ASB
option is activated)
ExTr (only if the flow
measurement + ASB
option is activated)
I:E
(only if the flow
measurement + ASB
option is activated)
Trigger time slot (only
in aPCV mode if the
pressure-controlled
ventilation modes
option is activated)
2
Unit
Description
ml
Tidal volume (breath volume)
1/min
Ventilation rate
mbar
Maximum inspiratory pressure
mbar
Positive end-expiratory pressure (CPAP)
Ventilation through the addition of
-
ambient air
-
Ventilation with concentrator oxygen
-
Ventilation with 100% oxygen
mbar
Pressure support (relative to the set PEEP)
-
Inspiratory trigger (three-level setting)
-
Expiratory trigger (three-level setting)
l/min
Inspiratory trigger (setting with units)
% Flow
Expiratory trigger (setting with units)
max
-
Inspiration to expiration ratio
-
Pressure increase time
% Te
Trigger time slot