4 Preparation and operation
Result

4.4.2 Inserting the device input filter

52
EN
MEDUMAT Standard
3. Perform a function check
check",
page 184).
The hygiene filter has been inserted.
1. Check the device input filter for external damage.
If necessary: Replace the device input filter.
Device may be damaged if a device input filter which has
already been pushed together is inserted in the filter
compartment!
On delivery, the filter cassette is inserted halfway into the device
input filter and is fixed in its position by a transport safety device.
If the filter cassette is pushed all the way into the device input filter
before insertion into the filter compartment of the device, the
function of the device input filter can no longer be guaranteed.
 Do not alter the state of device input filters on delivery.
 Do not push the filter cassette into the device input filter
completely by hand.
Transport
safety device
2. Remove the transport safety device from the device input filter.
3. Push the device input filter with the half-inserted filter cassette
into the filter compartment of the device.
In the process, the filter cassette is pushed all the way into the
device input filter.
4. Press the device input filter into the filter compartment until the
device input filter audibly clicks into place and sits flush with
the device.
2
(see "9.3 Performing a function
Filter cassette with filter fleece