11.6 Replacing the device input filter

Requirement
The device is switched off.
1. Pull the device input filter out of the filter compartment of the
2. Dispose of the device input filter along with the filter cassette
Device may be damaged if a device input filter which has
already been pushed together is inserted in the filter
compartment!
On delivery, the filter cassette is inserted halfway into the device
input filter and is fixed in its position by a transport safety device.
If the filter cassette is pushed all the way into the device input filter
before insertion into the filter compartment of the device, the
function of the device input filter can no longer be guaranteed.
 Do not alter the state of device input filters on delivery.
 Do not push the filter cassette into the device input filter
device.
(see "13.4 Hygiene filter/device input filter",
completely by hand.
MEDUMAT Standard
11 Maintenance
page 213).
2
EN
209