Parameters in the Physician menu ......... 8 Disable softstart generally ............ 9 Activate exchange mode ............ 10 Therapy data Save therapy data .............. 11 Read out therapy data ............11 Clear therapy data .............. 11 SOMNObalance (e) / SOMNOsoft 2 (e)
Alternatively, you can set the parameters using the SOMNOadjust remote setting system or WEINMANNsupport software. This document is not a complete set of instructions for use. Follow the instructions for use for the therapy device. Overview 2.1 Control panel SOMNObalance (e) / SOMNOsoft 2 (e)
Page 5
Scroll to next parameter of menu currently open. Open the info menu (with device on standby). Display current leakage (during therapy). Info key Save therapy data manually on the card. (Keep key depressed). Exit a menu. SOMNObalance (e) / SOMNOsoft 2 (e)
P min Pressure setting P max Speed of pressure rise softPAP stage: soft Parameter softPAP lock: SLOC Softstart pressure: P Lock info menu ILOC Exchange mode CHAN Device Time Date CLEAR Clear therapy data: SOMNObalance (e) / SOMNOsoft 2 (e)
The therapy device adopts the setting. 7. To exit the Physician menu, press the info key. Alternatively, wait 15 seconds. If you do not press any key for 15 seconds, the device automatically exits the Physician menu. SOMNObalance (e) / SOMNOsoft 2 (e)
Time hh:mm Date DD.MM.YYYY Clear therapy CLEAr data 3.3 Disable softstart generally If the therapy requires it, you can also disable softstart. To do so, set the softstart pressure of P to --.- hPa. SOMNObalance (e) / SOMNOsoft 2 (e)
This file can be read by WEINMANNsupport. • If exchange mode is activated, the therapy device cannot be switched on. • Exchange mode can be deactivated again via the Physician menu if it has been activated inadvertently. SOMNObalance (e) / SOMNOsoft 2 (e)
2. Press the humidifier key 3. Press the on/off key and then also the humidifier key until CLeAr flashes. The data are deleted as soon as the display stops flashing. This clearing step cannot be cancelled. SOMNObalance (e) / SOMNOsoft 2 (e)
Page 12
Weinmann Geräte für Medizin GmbH+Co. KG Postfach 540268 D-22502 Hamburg Kronsaalsweg 40 D-22525 Hamburg T: +49-(0)40-5 47 02-0 F: +49-(0)40-5 47 02-461 E: info@weinmann.de www.weinmann.de Center for Production, Logistics, Service Weinmann Geräte für Medizin GmbH+Co. KG Siebenstücken 14 D-24558 Henstedt-Ulzburg...