Weinmann SOMNObalance Instructions For Making Settings

Weinmann SOMNObalance Instructions For Making Settings

Cpap device with exhalation relief

Advertisement

SOMNObalance (e)
autoCPAP device with exhalation relief
SOMNOsoft 2 (e)
CPAP device with exhalation relief
Instructions for making settings

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SOMNObalance and is the answer not in the manual?

Questions and answers

Ислам
April 30, 2025

на дисплее выходит ошибка Int

Summary of Contents for Weinmann SOMNObalance

  • Page 1 SOMNObalance (e) autoCPAP device with exhalation relief SOMNOsoft 2 (e) CPAP device with exhalation relief Instructions for making settings...
  • Page 3: Table Of Contents

    Parameters in the Physician menu ......... 8 Disable softstart generally ............ 9 Activate exchange mode ............ 10 Therapy data Save therapy data .............. 11 Read out therapy data ............11 Clear therapy data .............. 11 SOMNObalance (e) / SOMNOsoft 2 (e)
  • Page 4: Introduction

    Alternatively, you can set the parameters using the SOMNOadjust remote setting system or WEINMANNsupport software. This document is not a complete set of instructions for use. Follow the instructions for use for the therapy device. Overview 2.1 Control panel SOMNObalance (e) / SOMNOsoft 2 (e)
  • Page 5 Scroll to next parameter of menu currently open. Open the info menu (with device on standby). Display current leakage (during therapy). Info key Save therapy data manually on the card. (Keep key depressed). Exit a menu. SOMNObalance (e) / SOMNOsoft 2 (e)
  • Page 6: Menu Structure

    P min Pressure setting P max Speed of pressure rise softPAP stage: soft Parameter softPAP lock: SLOC Softstart pressure: P Lock info menu ILOC Exchange mode CHAN Device Time Date CLEAR Clear therapy data: SOMNObalance (e) / SOMNOsoft 2 (e)
  • Page 7: Physician Menu

    The therapy device adopts the setting. 7. To exit the Physician menu, press the info key. Alternatively, wait 15 seconds. If you do not press any key for 15 seconds, the device automatically exits the Physician menu. SOMNObalance (e) / SOMNOsoft 2 (e)
  • Page 8: Parameters In The Physician Menu

    CPAP, APAP (SOMNObalance (e) only) CPAP SOMNObalance / SOMNOsoft 2: 4 hPa - 18 hPa If CPAP: CPAP pressure SOMNObalance e / SOMNOsoft 2e : 4 hPa - 20 hPa If APAP: lower 4 hPa-P pressure limit SOMNObalance / SOMNOsoft 2:...
  • Page 9: Disable Softstart Generally

    Time hh:mm Date DD.MM.YYYY Clear therapy CLEAr data 3.3 Disable softstart generally If the therapy requires it, you can also disable softstart. To do so, set the softstart pressure of P to --.- hPa. SOMNObalance (e) / SOMNOsoft 2 (e)
  • Page 10: Activate Exchange Mode

    This file can be read by WEINMANNsupport. • If exchange mode is activated, the therapy device cannot be switched on. • Exchange mode can be deactivated again via the Physician menu if it has been activated inadvertently. SOMNObalance (e) / SOMNOsoft 2 (e)
  • Page 11: Therapy Data

    2. Press the humidifier key 3. Press the on/off key and then also the humidifier key until CLeAr flashes. The data are deleted as soon as the display stops flashing. This clearing step cannot be cancelled. SOMNObalance (e) / SOMNOsoft 2 (e)
  • Page 12 Weinmann Geräte für Medizin GmbH+Co. KG Postfach 540268 D-22502 Hamburg Kronsaalsweg 40 D-22525 Hamburg T: +49-(0)40-5 47 02-0 F: +49-(0)40-5 47 02-461 E: info@weinmann.de www.weinmann.de Center for Production, Logistics, Service Weinmann Geräte für Medizin GmbH+Co. KG Siebenstücken 14 D-24558 Henstedt-Ulzburg...