Disinfecting The Reusable Measuring Hose System - Weinmann MEDUMAT Standard2 Instructions For Use Manual

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

11. If available: Connect the CO
to the measuring hose system connector.
12. If available: Seal the end of the reusable hose system closest to
the patient with a protective cap.
13. Place all the hoses and one of the connection lines in the hose
protection sleeve.
14. Close the Velcro fasteners in the hose protection sleeve around
the hoses and connection line.
15. Close the hose protection sleeve zipper.
Result
The reusable hose system is assembled.
8.7 Disinfecting the reusable measuring
hose system
The principle only applies to the following parts of the reusable
measuring hose system:
Pressure-measurement hose
PEEP control hose
Risk of injury due to incorrect disinfection of the measuring
hose system!
Rinsing the measuring hose system in the opposite direction to
that specified does not guarantee any bacteria reduction and may
injure the patient.
 Only rinse the pressure-measurement hose in the specified
direction.
8 Hygienic reprocessing
measuring hose with water filter
2
MEDUMAT Standard
2
EN
181