5 User menu
5.3.6 etCO
112
EN
MEDUMAT Standard
In this menu you can activate or deactivate apnea ventilation in the
CPAP, CPAP + ASB and Demand ventilation modes.
When apnea ventilation is activated, the device automatically
switches to IPPV mode once the set apnea time has elapsed.
If the BiLevel + ASB mode is unlocked in the operator menu and
activated, you can choose between the IPPV mode and the
BiLevel + ASB mode as the apnea ventilation mode in the operator
menu.
trend (only with capnography option)
2
5-7 etCO
trend submenu
2
If the capnography option is activated, the device offers the
possibility of a visualization of the etCO
be selected via the Views item in the user menu and shown on the
device's display.
The trend curve shows the patient's ventilation as a graphic. The
last value recorded appears on the far right of the trend curve.
2
value. The trend curve can
2