9 Function check
Result

9.3 Performing a function check

Requirement
184
EN
MEDUMAT Standard
5. Check the system for leaks
leaks",
page 192).
If necessary: Rectify any leaks in the system
leaks in the system",
The function check is ready.
You can perform the function check with the following test lungs:
Testing bag WM 1453
Testing bag WM 1454
EasyLung for WEINMANN Emergency WM 28625
Risk of injury from incorrect test lungs!
Test lungs other than those named here may not reliably detect
errors and thus distort the result of a function check. This can
injure the patient.
 Only use the test lungs named here.
Risk of injury due to a connection between the device and the
patient during the function check!
A connection between the device and the patient during the
function check can result in barotrauma and injury to the patient.
 Always disconnect any connection between the device and the
patient for the function check.
The device is disconnected from the patient.
A fully charged battery is inserted in the device.
The device input filter is inserted in the device.
The protective cap has been removed from the end of the
patient hose system.
The function check is ready
check",
page 183).
2
(see "9.6 Checking the system for
page 192).
(see "9.2 Preparing for the function
(see "9.7 Rectifying