Bluetooth - Weinmann MEDUMAT Standard2 Instructions For Use Manual

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

5 User menu
In this menu you can set the scale for the time axis of the etCO
2
trend. The following settings are possible: auto, 5 min., 10 min.,
30 min., 60 min.,120 min. At a time setting of 5 minutes or
10 minutes, the device records the determined value every
15 seconds. At a setting of 30, 60, or 120 minutes, an average
value is recorded every 30 seconds.
In the "auto" setting, the x axis is scaled automatically depending
on the duration of the application.

5.3.7 Bluetooth

®
(only with Bluetooth data
transmission option)
®
5-8 Bluetooth
submenu
If the Bluetooth data transmission option is unlocked and
activated in the operator menu
(see "4.14 Enabling options",
®
page 97), you can activate and deactivate the Bluetooth
®
connection here. If the Bluetooth
connection is activated, an
operation documentation system can connect to the device to
retrieve operating data.
If you have unlocked the Allow bluetooth pairing function in the
operator menu
(see "6.3.7 Device configuration",
page 125), the
MAC address of the device will also be displayed in this menu. You
2
can then pair MEDUMAT Standard
with an external data
documentation system during the application.
2
MEDUMAT Standard
EN
113

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wm 29500

Table of Contents