Pressure
pMax
CPAP / PEEP

7.3.4 CPAP + ASB mode

CPAP
1/Freq. (spontaneous)
The set value CPAP/PEEP is used to increase the pressure level of
respiration in order to raise the functional residual capacity (FRC)
of a spontaneously breathing patient. The patient is able to
breathe spontaneously without any restriction at the set pressure
level. The CPAP mode is used exclusively on patients with adequate
spontaneous respiration.
In principle, the pressure is set at the end of expiration (PEEP). The
set maximum pressure limitation (pMax) ensures the safety of the
patient.
Description
Abbreviation
Long form
Type
Requirement
7 Description of the modes
CPAP
1/Freq. (spontaneous)
CPAP + ASB
Continuous Positive Airway Pressure +
Assisted Spontaneous Breathing
Pressure-controlled
Flow measurement + ASB option is
activated
MEDUMAT Standard
Time
2
EN
143