7 Description of the modes

7.4.2 RSI mode

158
EN
MEDUMAT Standard
If you do not perform chest compressions for an extended period
of time in CPR CCSV mode, the device returns to IPPV back-up
ventilation after the time preset in the operator menu. As soon as
you recommence chest compressions, the device automatically
returns to CSSV ventilation.
The device always switches to IPPV mode upon exiting CPR mode.
If the capnography option is activated, you can set a pressure
gauge view or curve view in CPR mode
patient",
page 128).
Risk of injury from deactivated alarms in the additional
functions CPR and RSI!
Through deactivation in the operator menu, no alarms for flow
and CO
monitoring are output in the additional functions CPR
2
and RSI. This can injure the patient.
 Notify the user if the alarms in the additional functions CPR and
RSI have been deactivated.
 Observe the alarm limits during ventilation in the additional
functions CPR and RSI.
Description
Abbreviation
Long form
Type
Requirement
Ventilation parameters
Left-hand navigation knob
Central navigation knob
Right-hand navigation knob
The RSI mode supports you in the induction of anesthesia (TIVA). It
is used on all patients with an increased risk of a pulmonary
aspiration.
2
(see "6.3.8 Presets
RSI
Rapid Sequence Induction
Volume-controlled
None
Vt (only with RSI Manual)
-
pMax
Demand, manual, contin.
Emergency mode