5 User menu
5.3.8 NVG (Night Vision Goggles)
114
EN
MEDUMAT Standard
5-9 NVG submenu
Risk of injury from deactivated alarm light, deactivated audio
alarm output and darkened display in NVG mode!
The alarms are barely perceptible as a result of the deactivated
alarm light, the deactivated audio alarm output and the darkened
display in NVG mode. This can injure the patient.
Always monitor patients during ventilation.
Only use the NVG option in the military sector.
You can activate the NVG mode here. When the NVG mode is
activated, the device behaves as follows:
•
Alarm light deactivated
•
Acoustic alarm output for all alarms permanently deactivated
•
Line power and battery status indicators deactivated
•
Coloring of the display optimized for night vision devices
•
Display brightness reduced as per preset
configuration", page 125)
This submenu only appears if you activate the NVG option in the
operator menu
(see "6.3.9 Options",
only permitted for use in the military sector.
2
(see "6.3.7 Device
page 134). This option is