Risk of injury due to electric shock when connecting an
incorrect power supply to the line power!
The power supply contains a safety device to prevent electric
shock. The use of a non-original power supply may result in injury
to the user.
 Only operate the device on line power using the power supply
recommended by WEINMANN Emergency.
Caution
Risk of injury from touching the contacts in the battery
compartment and the patient at the same time!
The contacts in the battery compartment are live. Touching the
contacts and the patient at the same time can injure the user or
the patient.
 Do not touch the contacts in the battery compartment and the
patient at the same time.
Notice
Material damage due to prolonged storage of the battery
without recharging!
Storing the battery for a prolonged period of time without
recharging can result in the rapid shutdown of and irreparable
damage to the battery.
 When the battery is stored in the device without a power
connection: Charge battery every 3 months.
 If the battery is not stored in the device: Charge battery every
5 months.

2.1.3 How to use the patient hose system

Warning
Risk of injury due to contaminated or infected patient hose
system!
A patient hose system which is contaminated or infected as a
result of hygienic reprocessing not being performed or being
performed incorrectly may transmit contamination or infections to
the next patient and harm them.
 Do not reprocess disposable hose systems.
 Perform the hygienic reprocessing of reusable hose systems
correctly
system",
(see "8.4 Hygienic reprocessing of the reusable hose
page 168).
MEDUMAT Standard
2 Safety
2
EN
11