8.6.2 Assemble reusable hose system

Requirement
The reusable hose system is disassembled.
1. Assemble the patient valve.
2. If available: Secure the protective cap strap on the patient
3. Connect the patient valve to the ventilation hose.
When doing so, note:
the side of the PEEP control diaphragm labeled "TOP" must
face upward toward the control cover.
the arrow on the control cover must point toward the
patient.
valve:
Position 1
or
Position 2 (only with reusable hose systems with flow
measurement and CO
8 Hygienic reprocessing
1
measurement)
2
MEDUMAT Standard
2
2