Cpap Mode - Weinmann MEDUMAT Standard2 Instructions For Use Manual

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

7 Description of the modes

7.3.3 CPAP mode

142
EN
MEDUMAT Standard
The BiLevel + ASB mode is used for pressure-controlled ventilation
combined with free spontaneous respiration at pressure levels
pInsp and PEEP during the entire breathing cycle and for adjustable
pressure support at PEEP level. This mode is used on patients who
have no spontaneous respiration or on spontaneously breathing
patients. The patient can trigger a mandatory, pressure-controlled
mechanical breath during a predetermined trigger time slot. The
trigger time slot is 20% of the expiration time Te before the
anticipated mandatory mechanical breath. For the rest of the time,
the patient can breathe spontaneously or with the aid of pressure
support. Tidal volume and minute volume are determined by the
set pInsp, lung compliance and the set inspiration time T
Description
Abbreviation
Long form
Type
Requirement
Ventilation parameters
Left-hand navigation knob
Central navigation knob
Right-hand navigation knob
2
CPAP
Continuous Positive Airway Pressure
Pressure-controlled
None
-
-
PEEP
pMax
Emergency mode
.
i