8 Hygienic reprocessing
Result
8.5 Hygienic reprocessing of the
170
EN
MEDUMAT Standard
(3)
Only brush the outside of the part thoroughly using a normal
laboratory soft bottle brush. Do not use a bottle brush on the
inside in order to prevent damaging the sieves. Rinse thoroughly
with distilled water after cleaning.
(4)
Washing method according to the Robert Koch Institute (RKI)
and the Association for Applied Hygiene (VAH) at 70°C with Derval
SOLO (RKI) and with Ottalin PERACET)
2. Assemble reusable hose system
hose system",
page 177).
The reusable hose system has been hygienically reprocessed.
disposable hose system
1. If necessary: Release one of the following connection lines from
the hose clips:
•
Connection line of the MEDUtrigger
•
FlowCheck sensor connection line
•
FlowCheck sensor connection line with MEDUtrigger
2. Carry out hygienic reprocessing of the disposable hose system
as specified in the following table:
2
(see "8.6.2 Assemble reusable