2. Connect the open end of the nebulizer 1, 2, or 3 with the
3. Place the filter 7 (breathing system filter, bacteria filter, or a
4. When using the pneumatic drug nebulizer WM 15827:
Result
A nebulizer is connected.
4.5 Switching the device on
Requirement
•
•
•
1. Briefly press the On/Off button
filter 7 (breathing system filter, bacteria filter, or a combined
breathing system/bacteria filter).
combined breathing system/bacteria filter) on the patient hose
system's FlowCheck sensor 8.
Connect the oxygen hose to the inhalation outlet on the
pressure reducer.
When doing so, note: The nebulizer must be in a horizontal
position for sufficient nebulization to occur.
The device is disconnected from the patient.
A fully charged battery is inserted in the device.
The device is connected to the oxygen supply.
An automatic self-test starts, which runs through the following
sequence:
•
The alarm light flashes twice and two short test tones are
emitted
•
The start screen appears
4 Preparation and operation
.
MEDUMAT Standard
2
EN
65