9.1 Intervals

9.2 Preparing for the function check

Carry out a function check at regular intervals:
Part concerned
Device
Patient hose system (reusable hose
system)
1. Check battery status: The battery must be fully charged.
If necessary: Charge or replace the battery.
2. Check the following parts for external damage:
Device
Plug and cable
Patient hose system
Accessories
If necessary: Replace parts.
3. Check the patient valve of the patient hose system
"9.5 Testing the reusable hose system",
If necessary: Replace the patient hose system.
4. Check the oxygen level in the oxygen cylinder.
If necessary: Change the oxygen cylinder.
Interval
Before each use
After each hygienic reprocessing
After each repair
Before each use
After each hygienic reprocessing
After each disassembly
At least every 6 months
page 191).