9
Function check
9.1 Intervals
9.2 Preparing for the function check
Carry out a function check at regular intervals:
Part concerned
Device
Patient hose system (reusable hose
system)
1. Check battery status: The battery must be fully charged.
If necessary: Charge or replace the battery.
2. Check the following parts for external damage:
•
Device
•
Plug and cable
•
Patient hose system
•
Accessories
If necessary: Replace parts.
3. Check the patient valve of the patient hose system
"9.5 Testing the reusable hose system",
If necessary: Replace the patient hose system.
4. Check the oxygen level in the oxygen cylinder.
If necessary: Change the oxygen cylinder.
Interval
•
Before each use
•
After each hygienic reprocessing
•
After each repair
•
Before each use
•
After each hygienic reprocessing
•
After each disassembly
•
At least every 6 months
page 191).
MEDUMAT Standard
9 Function check
(see
2
EN
183