11 Maintenance

11.5 Replacing the hygiene filter

Requirement
Result
208
EN
MEDUMAT Standard
The device is switched off.
1. Wipe-disinfect the outside of the hygiene filter and the device.
2. Pull the hygiene filter out of the filter compartment of the
device.
3. Dispose of the hygiene filter along with the filter cassette
"13.4 Hygiene filter/device input filter",
4. With the filter side facing forwards, slide the hygiene filter into
the device's filter compartment until the hygiene filter is flush
with the device.
5. Perform a function check
check",
page 184).
The hygiene filter has been replaced.
2
page 213).
(see "9.3 Performing a function
(see