10 Alarms and error messages

10.2.2 Medium-priority alarm (yellow)

Alarm
Apnea mode active
Battery defective
Battery weak
Check the FlowCheck connection
line (only with flow
measurement + ASB option)
Check the FlowCheck sensor (only
with flow measurement + ASB
option)
CO
occlusion (only with
2
capnography option)
Compression rate
CCSV option)
Compression rate
CCSV option)
Disconnection MEDUtrigger
198
EN
MEDUMAT Standard
Cause
No inspiration since the set apnea
alarm time
Battery defective
Low battery status
FlowCheck sensor connection line/
FlowCheck sensor connection line
with MEDUtrigger not connected
correctly
FlowCheck sensor connection line/
FlowCheck sensor connection line
with MEDUtrigger defective
Software version of the FlowCheck
sensor connection line/FlowCheck
sensor connection line with
MEDUtrigger not compatible with
device
FlowCheck sensor not connected
correctly
FlowCheck sensor defective
Water filter blocked
CO
measuring hose blocked
2
(only with
Chest compression too fast
(only with
Chest compression too slow
MEDUtrigger removed from the
device during manual ventilation
2
Remedy
Check the condition of the
patient.
Select mandatory ventilation
mode.
Let the device run on battery
power without line power until it
switches off. Fully recharge battery
(see 4.3.2,
p. 47). If the device
continues to display the alarm:
Replace battery
(see 4.3.5,
Replace battery
(see 4.3.5,
Connect device to the line power
(see 4.2, p. 46)
and charge the
battery
(see 4.3.5,
p. 51).
Connect the FlowCheck sensor
connection line/FlowCheck sensor
connection line with MEDUtrigger
correctly.
Replace the FlowCheck sensor
connection line/FlowCheck sensor
connection line with MEDUtrigger
Perform a function check. The
device updates the software
version during the function check.
Connect the FlowCheck sensor
correctly.
Replace the FlowCheck sensor.
Replace water filter.
Replace CO
measuring hose.
2
Perform chest compression slower.
Perform chest compression faster.
Reconnect MEDUtrigger to the
device.
p. 51).
p. 51).