Performing Ventilation In Cpr Ippv Mode - Weinmann MEDUMAT Standard2 Instructions For Use Manual

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

4 Preparation and operation
Result

4.7.10 Performing ventilation in CPR IPPV mode

Requirement
Result
80
EN
MEDUMAT Standard
The device always switches to IPPV mode upon exiting CPR mode.
Ventilation is performed in CPR Manual mode.
Delay in treatment due to simultaneous metronome outputs
from the ventilator and the defibrillator!
If the ventilator is used together with a defibrillator (MEDUCORE
Standard) which can also emit a metronome sound, the
simultaneous metronome outputs from the defibrillator and the
ventilator may confuse the user and cause delays in treatment.
 Where the ventilator and defibrillator are used at the same
time, switch off the metronome sound of
MEDUMAT Standard
 Do not perform resuscitation without a metronome.
The device is switched on.
CCSV option is deactivated.
1. Briefly press the CPR button
Depending on the preset in the operator menu, the device
switches to CPR IPPV mode or CPR Manual mode (Operator
menu | Presets patient | CPR mode | Start mode).
2. If
is already set in the operator menu: Continue
ventilation in CPR IPPV mode
or
If Manual is already set in the operator menu: Switch to
using the right-hand navigation knob.
3. If the patient experiences spontaneous circulation again:
Cancel CPR IPPV mode using the CPR button
The device always switches to IPPV mode upon exiting CPR mode.
Ventilation is performed in CPR IPPV mode.
2
2
.
.
.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wm 29500

Table of Contents