4 Preparation and operation
Result

4.7.11 Performing ventilation in CPR CCSV mode

82
EN
MEDUMAT Standard
3. To start continuous ventilation again: Press the
Ventilation pauses.
When the countdown reaches zero, ventilation automatically
restarts.
Delay in treatment due to simultaneous metronome outputs
from the ventilator and the defibrillator!
If the ventilator is used together with a defibrillator, which can also
emit a metronome sound, the simultaneous metronome outputs
from both devices may confuse the user and cause delays in
treatment.
 Where the ventilator and defibrillator are used at the same
time, switch off the metronome sound of
MEDUMAT Standard
 Do not perform resuscitation without a metronome.
Treatment delays due to triggered alarms for spontaneously
breathing patients which do not apply in the this application
scenario!
In CPR CCSV mode, triggered etCO
and result in delays in treatment.
 Switch off etCO
menu.
2
2
.
alarms can confuse the user
2
alarms for CPR CCSV mode in the operator
2
field twice.