8.2 Intervals
Clean the device, components and accessories after every use (but
at least once a week).
8.3 Hygienic reprocessing of the device
Risk of injury due to reuse of disposable items!
Disposable items are intended for single use. Disposable items
which are reused may be contaminated and/or impaired in their
function and therefore cause injury to the patient.
Do not reuse disposable items.
Damage to the device caused by ingress of liquids!
The device is rated IP54 (splash-proof). This only takes effect when
the battery is located in the battery compartment. Ingress of
liquids may damage the device, components and accessories.
Do not immerse the device, components or accessories in
liquids.
Clean the battery compartment carefully so that no liquids
enter the device.
1. Disconnect the device from the patient.
2. Switch off the device
page 66).
3. If necessary: Disconnect the device from the line power.
4. Remove the battery.
5. Wipe-disinfect the outside of the device.
6. Remove the device input filter.
7. Disconnect the patient hose system from the device.
8 Hygienic reprocessing
(see "4.6 Switching the device off",
MEDUMAT Standard
2
EN