4.7.9 Performing ventilation in CPR Manual mode

Requirement
78
EN
MEDUMAT Standard
1. Assess CO
measurements diagnostically.
2
The CO
measurements and the respiratory rate of the patient are
2
monitored.
In CPR Manual mode, you determine the respiratory rate
administered yourself. Using MEDUtrigger, you manually trigger
individual mechanical breaths with the set tidal volume.
Delay in treatment due to simultaneous metronome outputs
from the ventilator and the defibrillator!
If the ventilator is used together with a defibrillator (MEDUCORE
Standard) which can also emit a metronome sound, the
simultaneous metronome outputs from the defibrillator and the
ventilator may confuse the user and cause delays in treatment.
 Where the ventilator and defibrillator are used at the same
time, switch off the metronome sound of
MEDUMAT Standard
 Do not perform resuscitation without a metronome.
The device is switched on.
MEDUtrigger is connected to the device.
MEDUtrigger is activated in the operator menu (Operator
menu | Options | MEDUtrigger).