4 Preparation and operation
4
Preparation and operation
4.1 Mounting the device
4.2 Connecting to a power supply
46
EN
MEDUMAT Standard
The device is mounted on a portable system as standard and is
ready for use. Observe the instructions for use of the portable
systems.
Loss of power due to combination of the device with an
incorrect power supply!
If you use a portable system which combines the
MEDUMAT Standard
power may occur in the devices in the event that these are used
with a 50 W power supply.
Use only the more powerful 100 W power supply when
combining the devices MEDUMAT Standard
Standard.
1. Check battery status
battery",
page 47).
2. If necessary: Charge battery
in the device",
page 47).
3. Slide full battery into the battery compartment until it clicks
into place.
2
2
and MEDUCORE Standard devices, a loss of
(see "4.3 Using the rechargeable
(see "4.3.2 Charging the battery
2
and MEDUCORE