11.4 Maintaining the reusable hose system

Requirement
The reusable hose system has been disassembled
"8.6.1 Disassembly of the reusable hose system",
1. Check all parts of the reusable hose system for external
2. Replace the PEEP control diaphragm and check valve
3. Assemble reusable hose system
4. Punch out the date at which the next maintenance is due on
5. Attach the service label to the end of the ventilation hose
6. Perform a function check
Result
The reusable hose system has been maintained and is ready for
use.
damage and complete labeling.
If necessary: Replace damaged or incorrectly labeled parts.
diaphragm (maintenance set WM 15779).
hose system",
page 177).
the service label (maintenance set WM 15779).
which is closest to the device.
check",
page 184).
(see "8.6.2 Assemble reusable
(see "9.3 Performing a function
MEDUMAT Standard
11 Maintenance
(see
page 172).
2
EN
207