Requirement
1. Briefly press the CPR button
2. If
3. If the patient experiences spontaneous circulation again: Stop
Result
Ventilation is performed in CPR CCSV mode.

Interrupting ventilation in CPR CCSV mode

During the analysis of the defibrillator, you can pause ventilation in
order to avoid ventilation artifacts in the analysis.
Requirement
1. Stop chest compressions.
Result
The ventilation is paused for as long as the chest compressions are
continued (up to the set back-up ventilation time at most).
The device is switched on.
The CCSV option is activated.
Depending on the preset in the operator menu, the device
switches to CPR CCSV mode or CPR Manual mode (Operator
menu | Presets patient | CPR mode | Start mode).
is already set in the operator menu: Continue
ventilation via chest compressions in CPR CCSV mode:
or
If Manual is already set in the operator menu: Switch to
using the right-hand navigation knob.
chest compressions.
or
Cancel CPR CCSV mode with the CPR button
The device is switched on
The CPR CCSV mode is set.