7 Description of the modes

7.4 Additional functions

7.4.1 CPR mode

156
EN
MEDUMAT Standard
The CCSV mode is a pressure-controlled ventilation mode
employed specially and exclusively for resuscitation ventilation
during continuing chest compression. The ventilation mode offers
you support during cardiopulmonary resuscitation (in accordance
with the resuscitation guidelines) by applying a defined pressure
into the lungs in time with every chest compression and switches
to expiration during the chest decompression phase.
The set inspiratory pressure corresponds to the maximum pressure
limit at the same time.
Risk of injury from deactivated alarms in the additional
functions CPR and RSI!
Through deactivation in the operator menu, no alarms for flow
and CO
monitoring are output in the additional functions CPR
2
and RSI. This can injure the patient.
 Notify the user if the alarms in the additional functions CPR and
RSI have been deactivated.
 Observe the alarm limits during ventilation in the additional
functions CPR and RSI.
Description
Abbreviation
Long form
Type
Requirement
2
CPR Manual
CPR IPPV
Cardiopulmonary Resuscitation
Volume-controlled
None
None
CPR CCSV
Pressure-
controlled
Flow
measurement
+ ASB option
is activated
CCSV option
is activated