Adjusting The Screen Brightness; Setting The Breathing Circuit; Setting The Hme Correction; Setting The Ventilation Mode For The Ventilation Category "Ventilation - Dräger Oxylog VE300 Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog VE300:
Table of Contents

Advertisement

– Bluetooth (G) (option)
– On, Off
– Screenshots (H) (option)
– On, Off
9.2

Adjusting the screen brightness

The screen brightness can be adjusted in a range from 25 % to 100 %.
To adjust the screen brightness:
● Touch the Screen brightness button (B), select the desired screen brightness
(25 %, 50 %, 75 % or 100 %) and confirm.
9.3

Setting the breathing circuit

The breathing circuit can be set for the use of disposable or reusable breathing
circuits.
To set the breathing circuit:
● Press the Breathing circuit button and select and confirm the breathing circuit
used (Disposable or Reusable).
9.4

Setting the HME correction

The temperature and moisture of the HME (heat and moisture exchanger) have an
effect on the flow measurement. The device can compensate for the presence of an
HME.
To set the HME correction:
● Touch the HME button, set the desired value (On or Off), and confirm the
setting.
9.5
Setting the ventilation mode for the ventilation
category "Ventilation (volume)"
Prerequisite: The Plus option is activated.
The ventilation mode for the ventilation category Ventilation (volume), which is
used as default in standby mode, can be set here. The options are VC-CMV and
VC-SIMV.
To set the ventilation mode:
● Touch the Ventilation (volume) button, set the desired mode, and confirm the
setting.
The ventilation mode can also be changed while ventilation is performed.
|
Instructions for use
Oxylog VE300 SW 1.n
Configuration
75

Hide quick links:

Advertisement

Table of Contents
loading