Download Print this page

Advertisement

Pocket Guide
Evita® V800 /
Evita® V600
Software version 1.n

Advertisement

loading

Summary of Contents for Dräger Evita V800

  • Page 1 Pocket Guide Evita® V800 / Evita® V600 Software version 1.n...
  • Page 2: Table Of Contents

    QUICK GUIDE EVITA® V800 / EVITA® V600 02 | Content Product description Operating concept Getting started Operation Alarms Monitoring The Evita® V800 / V600 Pocket Guide is not a replacement or substitute for the Instructions for Use. It is for informational purposes only . Any use of the device requires full understanding and strict observation of the Instruction for Use.
  • Page 3 | 03 Intended use The intensive care ventilator Evita is intended for the ventilation of adults, adolescents, children, infants, and neonates. This device provides mandatory ventilation modes and ventilation modes for supporting spontaneous breathing as well as ventilation monitoring.
  • Page 4: Product Description

    QUICK GUIDE EVITA® V800 / EVITA® V600 04 | Product description Intensive care Ventilator Display unit Ventilation unit Gas supply unit GS500 Power supply unit PS500 Trolley with 4 brakes Evita® V800...
  • Page 5 | 05 Product description Display unit, front Alarm bar Touch screen ON/OFF switch Indicator for power Rotary knob Alarm and batteries (touch turn confirm) silence 2 minutes...
  • Page 6 QUICK GUIDE EVITA® V800 / EVITA® V600 06 | Product description Display unit, rear COM ports USB ports HDMI port USB ports USB port LAN ports Swiveling of display unit...
  • Page 7 | 07 Product description Ventilation unit, front Flap Operation display of ventilation Expiration valve Expiratory flow sensor Inspiration valve Emergency air intake Nebulizer port Water trap...
  • Page 8: Operating Concept

    QUICK GUIDE EVITA® V800 / EVITA® V600 08 | Operating concept Screen, general structure Header Monitoring area Main menu Therapy...
  • Page 9 | 09 Operating concept Main screen Patient Ventilation mode / Power supply / Alarm messages Buttons to open the category Therapy type charge status - Alarms dialog - Views dialog indicator of - Trends/data dialog battery - Sensors dialog - Procedures dialog - System set up dialog Button to open Button to...
  • Page 10 QUICK GUIDE EVITA® V800 / EVITA® V600 10 | Operating concept Using the screen − Colours Alarms The following colors are used to identify the priority of alarms: Colour Alarm priority Action required High !!! Immediate action is necessary in order to avert an acute danger Yellow Medium !! Prompt action is necessary in order to avert a danger...
  • Page 11 | 11 Operating concept Using the screen − Colours Rotary knob If the device is turned on, the rotary knob is back lit with a color. Orange The selected function or setting must be confirmed by pressing the rotary knob. 10 seconds before the time during which the confirmation may be performed ends, the background lighting flashes.
  • Page 12: Getting Started

    QUICK GUIDE EVITA® V800 / EVITA® V600 12 | Getting started Turning on the device Prerequisites:  The accessories needed Ventilation unit, for operations on the patient left side: are connected. Main switch  The mains power supply and the gas supply are connected. ...
  • Page 13 | 13 Getting started System test Starting the system test 1. Touch the Start button in the System test line. 2. Confirm with the rotary knob. 3. Answer the questions in the information field by touching the Yes or No button. After a successful system test has been performed, a message appears for continuing the breathing...
  • Page 14 QUICK GUIDE EVITA® V800 / EVITA® V600 14 | Getting started Starting the therapy (for a new adult or paediatricpatient based on the patient height) 1. Touch the button for the relevant patient category. 2. Confirm with the rotary knob. 3.
  • Page 15 | 15 Getting started Therapy Quick start ventilation In an emergency situation, therapy can be started immediately without operational readiness being checked. 1. Touch the button Quick start ventilation. The Patient and therapy step is displayed. 2. Press the rotary knob. The therapy is started.
  • Page 16: Operation

    QUICK GUIDE EVITA® V800 / EVITA® V600 16 | Operation Changing the ventilation mode 1. Touch the relevant tab. 2. The colour of the tab turns orange (if necessary, adjust the ventilation parameter). 3. Confirm with the rotary knob. The color of the tab turns white. 1./2.
  • Page 17 | 17 Operation Advanced settings 1. Touch the button Advanced settings. The advanced settings of the selected ventilation mode are displayed and can be configured. 2. The current settings are displayed in an information field next to the button. Changing the therapy type during therapy 1.
  • Page 18 QUICK GUIDE EVITA® V800 / EVITA® V600 18 | Operation Non-invasive ventilation (NIV) In the Adult and Pediatric patient patient categories, all the ventilation modes can be selected. In the Neonate patient category, only the SPN-CPAP and PC-CMV ventilation modes can be selected. therapy Patient-side accessories such as oxygen masks or nasal cannulas...
  • Page 19 | 19 Operation Interrupting therapy 1. Touch the button Stop ventilation in the therapy bar. A pop-up window opens. 2. Touch the button Stop ventilation. 3. Confirm with the rotary knob. 4. The following message is displayed in the header bar: Standby mode activated 5.
  • Page 20: Alarms

    QUICK GUIDE EVITA® V800 / EVITA® V600 29 | Alarms Display of alarms – Optical alarm signals Alarm bar Alarm message field More alarms Blinking parameter field Display of alarms – Acoustic alarm signals The primary acoustic alarm system produces different alarm signals.
  • Page 21 | 21 Alarms Display of alarms – Confirming alarm messages The alarm logbook lists all alarms that are active and the ones which are no longer active. 1. Open the Alarm dialog page. 2. Touch the Reset all alarms button. 3.
  • Page 22: Monitoring

    QUICK GUIDE EVITA® V800 / EVITA® V600 22 | Monitoring Installing the CO sensor and CO cuvette 1. Insert the plug of the CO sensor into the socket at the rear of the ventilation unit. 2. Insert the cuvette into the patient port of the Y-piece.
  • Page 23 | 23 Monitoring monitoring Opening the dialog page Prerequisite: The CO sensor has warmed up. Time until availability: <10 s Time until specified accuracy is reached: <120s 1. Touch the Sensors button in the main menu bar. 2. Touch the CO sensor tab.
  • Page 24 Not all products, features, or services are for sale in all countries. Mentioned Trademarks are only registered in certain countries and not necessarily in the country in which this material is released. Go to www.draeger.com/trademarks to find the current status. CORPORATE HEADQUARTERS Drägerwerk AG &...

This manual is also suitable for:

Evita v600