Monitoring; Operating Data - Dräger Evita 2 dura Instructions For Use Manual

Intensive care ventilator
Hide thumbs Also See for Evita 2 dura:
Table of Contents

Advertisement

Technical Data

Monitoring

Operating data

Monitoring
Expiratory minute volume MV
Upper alarm limit alarm
Setting range
Lower alarm limit alarm
Setting range
Airway pressure Paw
Upper alarm limit alarm
Setting range
Lower alarm limit alarm
Insp. O
concentration FiO
2
Upper alarm limit alarm
Lower alarm limit alarm
Range
Insp. breathing gas temperature
Upper alarm limit alarm
Tachypnoea monitoring f
spn
Alarm
Adjustment range
Volume monitoring V
Ti
Lower alarm limit alarm
Upper alarm limit alarm
Adjustment range
Apnoea alarm time T
Apnoea
Alarm
Adjustment range
Operating data
Mains power connection
Current input
at 230 V
at 100 V
112
2
when MV exceeds the upper alarm limit.
0.1 to 1 L/min in 0.01 L/min increments
1 to 41 L/min in 0.1 L/min increments
when MV falls below the lower alarm limit.
0.01 to 1 L/min in 0.01 L/min increments
1 to 40 L/min in 0.1 L/min increments
when the "Paw high" value is exceeded.
10 to 100 mbar
when the value "PEEP +5 mbar" (coupled with the PEEP
set value) is not exceeded for at least 96 ms in 2 successive
ventilation strokes.
if FiO
exceeds the upper alarm limit for at least 20 seconds.
2
if FiO
falls below the lower alarm limit for at least 20 seconds.
2
both alarm limits are automatically allocated to the set value:
less than 60 Vol.%: ±4 Vol.%
60 Vol.% and over: ±6 Vol.%
when temperature reaches 40
(Evita 2 dura can also be used without temperature sensor
if the sensor is not connected on switching on).
when the spontaneous breathing frequency is exceeded
during spontaneous breathing.
5 to 120/min
if the set tidal volume V
(coupled with the set value V
T
not been supplied.
if the applied tidal volume exceeds the value of the alarm limit,
inspiration is interrupted and the expiration valve is opened.
21 to 4000 mL
if no breathing activity is detected.
5 to 60 s, adjustable in 1 second steps.
100 V to 240 V 50/60 Hz
max. 1.3 A
max. 3.2 A
o
C.
) has
T