Chapter 13 - Specifications

Operating Data

Operating voltage
Power input
Uninterruptible power supply
Auxiliary outlets
Connection for optional halogen lamp
Compressed gas supply at pipeline inlet
O
2
N
O
2
Air
Dew point
Oil content
Particles
Driving gas consumption
Noise emission in normal operation with
ventilation (breathing sound set to OFF)

Electrical Data

Protection class
Workstation
SpO
sensor
2
Electromagnetic compatibility (EMC)
210
100 to 127 VAC, 50-60 Hz, 12.4 A max.
200 W typically, max. 1.5 kW with power drawn from auxiliary outlets
Fully charged batteries: at least 30 minutes;
Up to 90 minutes, depending on ventilation parameters
(auxiliary outlets are not powered)
2 outlets with automatic circuit breakers rated at 4 A each,
1 outlet for desflurane vaporizer rated 2 A; combined current for all outlets
is 6A.
The outlets are isolated against mains to reduce leakage current.
12 V max. 10 W
39 psi to 100 psi at max. 78 L/min (including max. output flow)
39 psi to 100 psi at max. 18 L/min
39 psi to 100 psi at max. 18 L/min
>41°F below ambient temperature
3
<0.1 mg/m
Dust-free air (filtered with pores <1 µm)
None
<50 dB(A)
I, in accordance with IEC 601-1
Type BF
m
electrically isolated from protective conductor
Tested to IEC 601-1-2
Specifications
Apollo Operator's Manual