Operating data

Mains power connection
Current
at 230 V
at 100 V
Power consumption
Machine fuses
Range 100 V to 240 V
Protection class
Machine
CO
sensor (sensor connected)
2
Temperature sensor AWT 01
(sensor connected)
Gas supply
O
gauge pressure
2
O
connection thread
2
air gauge pressure
air connection thread
Gas consumption of control system
Output for pneumatic
medicament nebuliser
Automatic gas switch-over
Sound pressure level
(for free-field measurement over
a reflecting surface)
Dimensions (W x H x D)
Basic machine
Machine with trolley
Weight
Basic machine
Basic machine with trolley
incl. cabinet 8H

Machine outputs

Digital output
COM 1
100 V to 240 V
50/60 Hz
max. 1.3 A
max. 3.2 A
typically approx. 125 W
F 5 H 250 V IEC 127-2 (2x)
Class I
m
Type BF
m
Type BF
3 bar –10 % to 5.5 bar +10 %
at 60 L/min (peak flow 200 L/min)
M 12 x 1, female
3 bar –10 % to 5.5 bar +10 %
at 60 L/min (peak flow 200 L/min)
M 20 x 1.5, male
The gases must be dry and free from oil and dust.
Medical air or O
approx. 3.5 L/min
2
Medical air or O
2
max. 2 bar
max. 10 L/min
if one gas fails (inlet pressure <1.5 bar),
the device switches to the other gas.
max. 47 dB (A)
530 x 290 x 450 mm
580 x 1335 x 660 mm
approx. 27 kg
approx. 69 kg
Output and reception via an RS 232 C interface
LUST protocol
Baudrate: 1200, 2400, 4800, 9600, 19200 baud
Data bits: 7
Parity: even
Stop bits: 1
MEDIBUS protocol
Baudrate: 1200, 2400, 4800, 9600, 19200 baud
Data bits: 8
Partiy: even, odd, no
Technical Data
157