Operating Data - Dräger Oxylog 2000 Operating Instructions Manual

Emergency & transport ventilator
Hide thumbs Also See for Oxylog 2000:
Table of Contents

Advertisement

Technical Data

Operating Data

Operating Data
Gas supply
Supply gas
Quality of the supply gas
Supply
Supply pressure
Gas cylinders and pressure regulators
Pressure regulators
Connection
Gas consumption for internal control
with »Air Mix« selected
with »No Air Mix« selected
Typical time of operation
at a minute ventilation of 10 L/min
D-type O
cylinder
2
E-type O
cylinder
2
Electrical supply
Input voltage for Oxylog 2000
Connection for
external power supply
12 V / 24 V / 28 V DC
54
Oxygen for medical use, mixture of gases,
medical air in emergencies
WARNING !
Always use medical grade oxygen or air that is
dry and free from dust and oil. Contaminated
gas may cause ventilator malfunction.
From a pipeline system
or
from medical gas cylinders
40 to 88 psi ( 2.7 to 6.0 bar) at 80 L/min
WARNING !
Always use gas cylinders and pressure
regulators that are officially approved and
comply with all applicable safety regulations.
must have a safety relief valve on the output side to limit
output pressure to approx. 74 psi (5 bar) in the event of a defect.
DISS O
(male),
2
may also be used for medical air in an emergency
Approximately 1.0 L/min
Approximately 50 % of the effective minute ventilation
Approximately 100 % of the effective minute ventilation
Approximately 32 minutes without mixing (No Air Mix)
Approximately 64 minutes with mixing (Air Mix)
Approximately 56 minutes without mixing (No Air Mix)
Approximately 112 minutes with mixing (Air Mix)
12 V ±1 V DC
with DC/DC converter