Configuring in Standby Mode
Setting Default Values
CMV Default Ventilation Settings
Select and confirm the line »CMV default settings«
with dial knob.
The menu keys for CMV are displayed:
Display (example):
To modify settings:
Press the respective menu key.
Set and confirm with dial knob.
To return to »CMV default settings«:
Press dial knob.
PCV Default Ventilation Settings
Select and confirm the line »PCV default settings«
with dial knob.
The menu keys for PCV are displayed:
Display (example):
To modify settings*:
Press the respective menu key.
Set and confirm with dial knob.
To return to the »PCV default settings« menu:
Press dial knob.
_____________
*
Maximum default setting for Pmax: 30 cmH
100
Standby / Configuration
MAC-config.
MAC display
on
MAC correction
on
age
Adjust default settings or confirm
Standby / Configuration
MAC-config.
MAC display
on
MAC correction
on
age
Adjust default settings or confirm
Fresh Gas O
O %
2
100
O.
2
defaults
calibration
pls. tone
O -sensor
2
alarm sound
21 vol.%
off
scale
gas
parameter
sensor
off
record
more
interface
36
alarm limits
curves
basic configurations
CMV default settings
PCV default settings
Fresh Gas Flow Control
P
V
Freq.
max
T
cmH O
mL
bpm
2
25
600 12
calibration
defaults
pls. tone
O -sensor
2
alarm sound
21 vol.%
off
scale
gas
parameter
sensor
off
record
more
36
interface
alarm limits
curves
basic configurations
CMV default settings
PCV default settings
Fresh Gas Flow Control
! Select default carrier gas using hard key.
2
L/min
2.00
0 1 2 3 4
5
6 7 8 9
1 2 3
4
5 6 7 8 9
adult
child
T
:T
T :T
PEEP
IP I
I
E
%
cmH O
2
1:2.0
10
0
0 1 2 3 4
5
6 7 8 9
1 2 3
4
5 6 7 8 9
adult
child