Emergency start

Use only in urgent cases if there is no time for
the self test!
1 Switch on the workstation (A).
2 Check that both vaporizers are closed.
3 Set the safety knob (B) of the O
livery to the required O
12 L/min.
4 Start manual ventilation.
Wait for the internal loading of the software and
the testing of the electronics. The check list is
displayed after about 35 seconds.
Instructions for use Primus Infinity Empowered SW 4.5n
B
A
emergency de-
2
flow, between 0 and
2
Primus
Check List
5 Press the soft key cancel test (C).
The device only runs through a minimal self test
for about 10 seconds. Manual ventilation is in-
terrupted during this phase. Spontaneous
breathing can continue.
Primus Infinity Empowered is ready for operation
about 1 minute after initiating. The O
completely calibrated after about 5 minutes.
The leak and compliance test is not performed. The
accuracy levels specified in the chapter "Technical
data" on page 247 cannot be guaranteed.
NOTE
To prevent abuse of this feature, the self test can
only be canceled ten times in succession.
After ten cancellations, the self test cannot be can-
celed the next time that Primus Infinity Empowered
is started and a complete self test must be carried
out.