Set startup settings

Range of the settings:
Parameter
Range
Trigger
0 (OFF) 3 to 15 L/min
PEEP
0 to 20 mbar
I:E
3:1 to 1:4
Tinsp
0.2 to 10.0 s
Tplat %
0 to 50%
ΔASB
0 to 35 mbar
Ramp
SLOW, STANDARD, FAST
Pinsp
3 to 55 mbar
O
-Flow
0 to 15 L/min
2
NIV
ON, OFF
Tapn
0 (OFF) to 60 s
VTapn
50 to 2000 mL
fapn
12 to 60 bpm
MV-high
2.0 to 41 L/min
MV-low
0.5 to 40 L/min
fspn-high
10 to 100 bpm
Loudness
1/4 to 4/4
Brightness
1/4 to 4/4
The default settings for the parameters are
displayed on the screen when the ventilator is
switched ON. The settings can be adjusted.
Instructions for Use Oxylog 3000 SW 1.n
Advance to the second page:
1 Select the line Page, confirm and turn rotary
knob.
To restore the manufacturer's defaults:
2 Select and confirm line Set factory default.
Select hose type
The type of ventilation hose (reusable or
disposable hose system) can be configured.
The gas outlet connectors for the flow measuring
hoses must correspond to the hose type system
selected.
Configuration
81