Fgt Menu - Anritsu MS2024B Manual

Vector network analyzer for anritsu rf and microwave handheld instruments
Hide thumbs Also See for MS2024B:
Table of Contents

Advertisement

6-3 Domain Setup Menus

FGT Menu

Preliminary Key Sequence: Measure > Domain Selection > FGT
or
Preliminary Key Sequence: Freq/Time/Dist > Domain Setup > Setup FGT
Key Sequence: Freq/Time/Dist
Start Freq: Press this soft key to set the start frequency in units of Hz, kHz,
MHz, or GHz. If you enter a start frequency that is higher than the
FGT
currently-set stop frequency, and if CAL = On, then the start frequency is set
Start Freq
to the same value as the stop frequency. If you enter a start frequency that is
# kHz
higher than the currently-set stop frequency, and if CAL = Off, then both start
and stop frequency are set to the new frequency.
Stop Freq
Stop Dist: Press this soft key to set the stop frequency in units of Hz, kHz,
# GHz
MHz, or GHz. If you enter a stop frequency that is lower than the currently-set
start frequency, and if CAL = On, then the stop frequency is set to the same
value as the start frequency. If you enter a stop frequency that is lower than
Center Freq
# GHz
the currently-set start frequency, and if CAL = Off, then both stop and start
frequency are set to the new frequency.
Center Freq: Press this soft key to set the center frequency in units of Hz,
Span
# GHz
kHz, MHz, or GHz. The center frequency can be set with the arrow keys, the
rotary knob, or the number keypad. When using the number keypad, the
menu displays soft keys with Hz, kHz, MHz, and GHz units. Pressing the
Gate
Enter key has the same effect as pressing the MHz soft key.
Span: Press this soft key to set the span in units of Hz, kHz, MHz, or GHz.
The span setting determines the frequency range over which the instrument
sweeps. The span may be set from 0 (zero) to the maximum frequency of the
instrument.
Gate: Press this soft key to display the
Domain Setup
FGT
Domain Setup: Press this soft key to return to the
The message on the key face reflects the setup feature (Distance) that is in
use.
Figure 6-13. FGT Menu
6-14
PN: 10580-00289 Rev. K
VNA View Menus
"Gate
Menu".
"Setup Domain
Vector Network Analyzer MG
Menu".

Advertisement

Table of Contents
loading

Table of Contents