6-8 System Menus

Application Options Menu (SPA Mode) LMR Master

The display of this menu varies based on the current instrument measurement mode. The
menu shown here is specific to the Spectrum Analyzer application of the S412E LMR Master.
Key Sequence: Shift, System (8) > Application Options
Options
Impedance
50 Ohm 75 Ohm Other
Bias Tee
Back
Bias Tee
Bias Tee
Off
On
Bias Tee Voltage
15.0 V
Current
Low
High
Back
Figure 6-65. Application Options Menu (SPA Mode) LMR Master
6-66
SPA Mode selected
Impedance
50 Ohm 75 Ohm Other: Press this soft key to toggle the
impedance setting to 50 Homs, 75 Ohms, or Other impedance value.
Selecting 75 Ohm selects the 7.5 dB loss of the Anritsu 12N50-75B
adapter. For other adapters, select Other and enter the appropriate loss
value.
Bias Tee (Option 10 required): Press this soft key to display the Bias
Tee menu.
Back: Press this soft key to return to the
Bias Tee
Off On: Press this soft key to toggle the variable power supply Off
and On.
Bias Tee Voltage: Use this soft key to set the power supply voltage.
Current
Low High: Press this soft key to toggle the current between Low and
High.
Back: Press this soft key to return to the Application Options menu.
PN: 10580-00289 Rev. K
VNA View Menus
"System
Menu".