Bias Tee Menus (Ms20Xxc, Ms2026B, And Ms2028B) - Anritsu MS2024B Manual

Vector network analyzer for anritsu rf and microwave handheld instruments
Hide thumbs Also See for MS2024B:
Table of Contents

Advertisement

Bias Tee, Option 10

10-5 Bias Tee Menus (MS20xxC, MS2026B, and MS2028B)

To access the Bias Tee menu, press the Sweep function hard key (or press the Shift key then
the Sweep (3) key). Press the Configure Ports soft key, and then press the Bias Tee Setup soft
key to open the Bias Tee Setup menu. Press the Bias Tee soft key to select Internal, External, or
Off.
Sweep
Run/Hold
Run
Hold
Sweep Type
Single
Cont
Data Points
###
IFBW
# Hz
Sweep Averaging
#
Configure Ports
RF Pwr in Hold
On
Off
Figure 10-6. Bias Tee Menu Group (MS20xxC, MS2026B, and MS2028B)
For additional information about the Sweep menu, refer to
on page 6-53
and to
Vector Network Analyzer MG
10-5 Bias Tee Menus (MS20xxC, MS2026B, and MS2028B)
Configure Ports
Auto Reference
Plane Extension
Port 1 Ref Plane Length
# mm
Ext
Port 2 Ref Plane Length
# mm
DUT Line Type
Coax
Waveguide
Propagation Velocity
#
Bias Tee Setup
Source Power
High
Back
"Sweep Menu — MS20xxB and S412E" on page
PN: 10580-00289 Rev. K
Bias Tee Setup
Bias Tee
Off
Int Port Selection
1
2
Int Voltage P1
##.# V
Int Current Limit P1
## mA
Int Voltage P2
##.# V
Int Current Limit P2
## mA
Back
"Sweep Menu — MS20xxC"
6-54.
Bias Tee
Off
External
Internal
Back
10-7

Advertisement

Table of Contents
loading

Table of Contents