Dut Port Setup Menu (Waveguide) - Anritsu MS2024B Manual

Vector network analyzer for anritsu rf and microwave handheld instruments
Hide thumbs Also See for MS2024B:
Table of Contents

Advertisement

VNA View Menus

DUT Port Setup Menu (Waveguide)

Key Sequence: Shift, Calibrate (2) > DUT Port Setup
WG DUT Port 1: Press this soft key to open the "DUT Connector Selector"
List Box for Port 1. Choose the connector type that matches the DUT
DUT Port Setup
connection on Port 1. This indicates that you are using the matching
WG DUT Port 1
calibration kit, and the VNA Master displays the corresponding calibration
WG11A/WR229/R40
coefficients that will be used during the calibration procedure.
WGDUT Port 2
WG11A/WR229/R40
Setup User–Defined
SOLT(#)–USER #
WG DUT Port 2: Press this soft key to open the "DUT Connector Selector"
List Box for Port 2. Choose the connector type that matches the DUT
connection on Port 2. This indicates that you are using the matching
Back
calibration kit, and the VNA Master displays the corresponding calibration
coefficients that will be used during the calibration procedure.
Setup User-Defined: Press this soft key to open the "User-Defined Cal Kit
Selector" List Box
user-defined cal kit displays the cal kit definition menu
menu, you may enter the calibration coefficients for their specific calibration
kit.
Back: Press this soft key to return to the
Figure 6-32. DUT Port Setup Menu
Vector Network Analyzer MG
Examples:
WG11A
WG12
WG20
SSLT(1) – USER 1
SSLT(4) – USER 4
(Figure
6-34). Format: SSST(#) – USER #. Selecting a
PN: 10580-00289 Rev. K
6-4 Calibration Menus
(Figure
"Calibration
Menu".
6-33). In this
6-33