Field View Menus

Sweep Menu

Run/Hold
Run Hold: Press this soft key to toggle the sweep setting to Run or Hold.
Sweep
Sweep Type
Run/Hold
Single Cont: Press this soft key to toggle the sweep setting to Single or
Run
Hold
Continuous (Cont).
Data Points: Press this soft key to set the number of data points to be
Sweep Type
included in the sweep. You can set any number of data points from 2 up
Single
Cont
to 4001. The larger the number of data points, the slower sweep speed. In a
DTF measurement for example, the larger the number of data points, the
Data Points
longer the maximum distance, but at the expense of a slower sweep speed.
##
Avg/Smooth: Press this soft key to display the
on page
Avg/Smooth
Propagation Velocity: Press this soft key to set the propagation velocity for
the cable under test. This value is set automatically when a cable type is
selected for a distance measurement (refer to
Propagation Velocity
on page
#
Output Power: Press this soft key to display the
on page
Output Power
RF Immunity
High
High Low: Press this soft key to toggle the RF Immunity setting to High or
RF Immunity
Low.
RF Pwr in Hold
High
Low
On Off: Press this soft key to toggle On and Off the condition of the RF
RF Pwr in Hold
power transmitted from Port 1 and Port 2 when the instrument is put in Hold
mode. The RF power can be set either to stay on or to be turned off during
On
Off
Hold. The default setting is for the RF to stay On during Hold, which helps to
stabilize the instrument temperature.
Figure 7-12. Field View Sweep Menu
Vector Network Analyzer MG
7-14.
7-6).
7-14.
PN: 10580-00289 Rev. K
7-6 Sweep Menus