VNA View Menus

Gate Menu

Key Sequence: Measure > Domain Selection > Gate
Gate Function
Off Display On: Press this soft key to toggle the Gate Function status to
Gate
Off, Display, or On.
Gate Function
Off
Display
On
Start Gate
# ns
Stop Gate
Start Gate: Press this soft key to set the start Gate value. The values will be
# ns
in units of time if the active trace is in the Time Domain, or in units of distance
if the active trace is in the Distance Domain. Use the arrow keys, the rotary
Center Gate
knob, or the number keypad to set a Gate value. When using the number
# ns
keypad, the menu displays soft keys with seconds (s), milliseconds (ms),
microseconds (µs), nanoseconds (ns), or picoseconds (ps), if in the Time
Span Gate
Domain. Otherwise, it displays meters (m), centimeters (cm), or
# ns
millimeters (mm) as units, if in Distance Domain. Pressing the Enter key has
the same effect as pressing the seconds (s) soft key or the meter (m) soft key.
Gate Notch
If the
On
Off
feet rather than meters, then the distance settings are in feet (ft) only, and
pressing the Enter key sets the units to feet.
Gate Shape
Stop Gate: Press this soft key to set the stop Gate in either time or distance
Minimum
units, depending on whether the active trace is in the Time Domain or the
Distance Domain. In time domain, the units are in seconds (s),
milliseconds (ms), microseconds (µs), nanoseconds (ns), or
Back
picoseconds (ps). In distance domain, the units are in meters (m),
centimeters (cm), or millimeters (mm). If the
Mode)"
settings are in feet (ft) only.
Figure 6-14. Gate Menu
This menu description is continued on the next page.
Vector Network Analyzer MG
When set to Off, the gate is not displayed on the screen nor applied to
the data.
When set to Display, the gate shape is displayed on the screen but the
gate filtering is not applied to the data.
When set to On, the gate shape is displayed on the screen, and the
gate filtering is applied to the data.
"Application Options Menu (VNA Mode)"
menu is used to set units in feet rather than meters, then the distance
PN: 10580-00289 Rev. K
6-3 Domain Setup Menus
menu is used to set units in
"Application Options Menu (VNA
6-15