Bias Tee Setup Menu - Anritsu MS2024B Manual

Vector network analyzer for anritsu rf and microwave handheld instruments
Hide thumbs Also See for MS2024B:
Table of Contents

Advertisement

6-7 Measurement Menus

Bias Tee Setup Menu

Access to this menu is from the Bias Tee Setup soft key in the
Bias Tee: Press this soft key to open the
select External, Internal, or Off.
Bias Tee Setup
Int Port Selection
Bias Tee
1 2: Press this soft key to toggle the internal port selection to Port 1 or
Off
Port 2.
Int voltage P1: Press this soft key to set the internal bias tee voltage that is
Int Port Selection
directed onto the center conductor of port 1. The available range is from
1
2
12.0 V to 32.0 V in increments of 0.1 V. Use the arrow keys, the rotary knob,
or the number keypad to change the setting. When using the number keypad,
Int Voltage P1
press the soft key for voltage units (V), or press the Enter key. Press the Esc
##.# V
key to exit without changing the setting.
Int Current Limit P1: Press this soft key to set the internal bias tee current
Int Current Limit P1
limit for the voltage that is set at Port 1. The available range is from 0 mA to
## mA
450 mA in steps of 1 mA. This current limit sets the trip point for the bias tee
for this port. When using the number keypad, the soft key menu displays
Int Voltage P2
2 choices for units: A or mA. Use the arrow keys, the rotary knob, or the
##.# V
number keypad to change the setting. When using the number keypad, press
the soft key for current units (A or mA), or press the Enter key to use mA.
Press the Esc key to exit without changing the setting.
Int Current Limit P2
## mA
Int voltage P2: Press this soft key to set the internal bias tee voltage that is
directed onto the center conductor of Port 2. The available range is from
12.0 V to 32.0 V in increments of 0.1 . Use the arrow keys, the rotary knob, or
the number keypad to change the setting. When using the number keypad,
press the soft key for voltage units (V), or press the Enter key. Press the Esc
key to exit without changing the setting.
Back
Int Current Limit P2: Press this soft key to set the internal bias tee current
limit for the voltage that is set at Port 2. The available range is from 0 mA to
450 mA in steps of 1 mA. This current limit sets the trip point for the bias tee
for this port. When using the number keypad, the soft key menu displays
2 choices for units: A or mA. Use the arrow keys, the rotary knob, or the
number keypad to change the setting. When using the number keypad, press
the soft key for current units (A or mA), or press the Enter key to use mA.
Press the Esc key to exit without changing the setting.
Back: Press this soft key to return to the Configure Ports menu.
Figure 6-54. Bias Tee Setup Menu
6-56
PN: 10580-00289 Rev. K
VNA View Menus
"Configure Ports
"Bias Tee Menu (MS20xxC only)"
Vector Network Analyzer MG
Menu".
to