6-7 Measurement Menus
Source Power Menu (MS20xxC only)
Access to this menu is from the Source Power soft key in the
Source Power
Low: Press this soft key to set the Source Power to Low. This setting is for
Low
amplified devices.
–15 to –25dBm
High: Press this soft key to set the Source Power to High. This setting is for
High
3 to –3dBm
passive devices.
Back: Press this soft key to return to the
Back
Figure 6-56. Source Power Menu
Source Power Menu (MS20xxB and S412E only)
Access to this menu is from the Source Power soft key in the
Source Power
Low: Press this soft key to set the Source Power to Low, which is
Low
approximately –25 dBm. This setting is for amplified devices.
~ –25 dBm
Default: Press this soft key to set the Source Power to Default, which is
Default
~ –5 dBm
approximately –5 dBm. This setting is for passive devices.
High: Press this soft key to set the Source Power to High, which is
High
~ 0 dBm
approximately –0 dBm. Use this setting to achieve maximum dynamic range
when measuring high-loss devices.
Back
Back: Press this soft key to return to the
Figure 6-57. Source Power Menu