11-6 Calibration Correction
1
1 The VNA Master
2 Optional Test Port Cable
3 Optional Adapter for Through Connection
4 OSL (Open, Short, Load) Precision Calibration Components
Figure 11-3. 2-Port Calibration, using MS2028C
Figure 11-3
test port cables, and if you have sufficient calibration components, then using male
Note
OSL standards on one port and female OSL standards on the other port allows you
to connect the cable ends together without the adapter. This becomes more
important at higher frequencies.
11-8
3
2
shows an optional adapter for the Through connection. If you are using
PN: 10580-00289 Rev. K
Vector Voltmeter, Option 15
4
OPEN
OPEN
SHORT
LOAD
SHORT
LOAD
2
Vector Network Analyzer MG