6-4 Calibration Menus
VNA View Menus
Calibration Methods
Figure 6-22. Calibration Method List
The Reciprocal is also known as an Unknown Thru. If you do not know the length of the Thru
in your setup, and if it is a reciprocal device, then you can choose one of the bottom three
calibration methods shown in
Figure
6-22. Set the Line Length to zero, and the vector network
analyzer will automatically measure (calculate) the length of the Thru.
6-24
PN: 10580-00289 Rev. K
Vector Network Analyzer MG