Table Menu - Anritsu MS2024B Manual

Vector network analyzer for anritsu rf and microwave handheld instruments
Hide thumbs Also See for MS2024B:
Table of Contents

Advertisement

Vector Voltmeter, Option 15
Continuous Wave (CW) Measurements
In Continuous Wave, the meter displays the selected frequency and indicates if calibration is
On or Off for that specific frequency. The instrument must be calibrated for each chosen
frequency.
The return measurement format units are displayed near the center of the Vector Voltmeter
window. For dB format, the units are dB and deg (degrees). For VSWR format, the voltage
standing wave ratio values are unitless. For Imped (impedance) format, the units are Ohms
and jOhms.
The measurement type and format are displayed at the bottom of the Vector Voltmeter
window.

Table Menu

CW Frequency: Press this soft key to set the desired measurement
frequency. Enter the desired frequency using the keypad, the arrow keys, or
Table
the rotary knob. If entering a frequency using the keypad, then the soft key
CW Frequency
labels change to GHz, MHz, kHz, and Hz. Press the appropriate units key.
# GHz
Pressing the Enter key has the same affect as using the MHz soft key.
Measurement Type
Measurement type
Return Insertion: Press the Measurement Type soft key to toggle between
Return
Insertion
return and insertion.
Select Cable: Press this soft key to open the Select Cable List Box.
Select Cable
One of the twelve cable numbers is underlined to indicate which cable is
#
selected. Change to a different cable by pressing the Select Cable soft key.
The data for the selected cable is highlighted in that cable row within the
Save
displayed table.
New
Reference
Save New Reference: Press this soft key to save the current measurement
as the reference measurement. Entering a reference puts the Vector
Clear
Voltmeter in relative measurement mode.
Reference
Clear Reference: Press this soft key to remove the reference measurement
from memory. The Vector Voltmeter exits the relative measurement mode.
Cal Port
Port_1
Port_2
Cal Port
Port_1 Port_2: Press this soft key to set the transmission port for
Source Power
subsequent measurements. For making S
Port 1. For making S
High
is displayed only with the MS20xxC VNA Master and is not relevant to the
MS20xxB VNA Master.
IFBW
Source Power: Press this soft key to open the
# kHz
(MS20xxB and S412E only)" on page
mode is the same as in VNA mode.
IFBW: Press this soft key to open a list box and set the Intermediate
Frequency bandwidth. The default is 10 kHz. Select 10 Hz for the maximum
dynamic range, and select 100 kHz for the maximum speed.
Figure 11-9. Table Menu
Vector Network Analyzer MG
or S
measurements, set this to Port 2. This soft key
22
12
PN: 10580-00289 Rev. K
11-9 Vector Voltmeter Menus
or S
measurements, set this to
11
21
"Source Power Menu
6-58. The Source Power menu in VVM
11-15