Bias Tee Setup Menu - Anritsu MS2024B Manual

Vector network analyzer for anritsu rf and microwave handheld instruments
Hide thumbs Also See for MS2024B:
Table of Contents

Advertisement

Bias Tee, Option 10

Bias Tee Setup Menu

Bias Tee Setup
Bias Tee (On/Off): Press this soft key to open the
On or Off the Bias Tee voltage.
Bias Tee
Off
Int voltage P2: Press this soft key to set the internal bias tee voltage that is
directed onto the center conductor of Port 2. The available range is from
12.0 V to 32.0 V in increments of 0.1 . Use the arrow keys, the rotary knob, or
the number keypad to change the setting. When using the number keypad,
Int Voltage P2
press the soft key for voltage units (V), or press the Enter key. Press the Esc
key to exit without changing the setting.
##.# V
Int Current Limit P2: Press this soft key to set the internal bias tee current
Int Current Limit P2
limit for the voltage that is set at Port 2. The available range is from 0 mA to
450 mA in steps of 1 mA. This current limit sets the trip point for the bias tee
## mA
for this port. When using the number keypad, the soft key menu displays
2 choices for units: A or mA. Use the arrow keys, the rotary knob, or the
number keypad to change the setting. When using the number keypad, press
the soft key for current units (A or mA), or press the Enter key to use mA.
Press the Esc key to exit without changing the setting.
Back
Back: Press this soft key to the Sweep menu.
Figure 10-10.Bias Tee Setup Menu – MS20xxB and S412E
Bias Tee Menu
Bias Tee
Off: Press this soft key to turn Off the Bias Tee function.
Off
On: Press this soft key to turn ON the Bias Tee function.
On
Back: Press this soft key to return to the
changing the current Bias Tee setting.
Back
Figure 10-11. Bias Tee Menu – MS20xxB and S412E
Vector Network Analyzer MG
10-6 Bias Tee Menus (Field Measurements View)
PN: 10580-00289 Rev. K
"Bias Tee Menu"
"Bias Tee Setup Menu"
to turn
without
10-11