2-Port Phase Measurement - Anritsu MS2024B Manual

Vector network analyzer for anritsu rf and microwave handheld instruments
Hide thumbs Also See for MS2024B:
Table of Contents

Advertisement

Field Measurements
The example in this figure may not match the display on your instrument.
Figure 5-9.
Return Loss, 1-Port Phase Measurement

2-Port Phase Measurement

The following example compares the phase of two cables using a 2-port phase measurement.
Figure 5-10
shows a typical 2-Port Phase measurement (Insertion Loss). This screen display
may not match the display on your instrument.
Procedure
1. Press the Measure function hard key and then press the Measurement Type soft key and
select 2-Port Phase.
2. Press the Freq/Dist function hard key and set the Start Frequency and Stop Frequency.
3. Press the Sweep function hard key and then the Data Points soft key to set the number
of data points (the larger the number of data points, the longer the maximum distance,
at the expense of a slower sweep speed).
4. Press the Shift key, then the Calibrate (2) key.
5. Press the Start Cal soft key and perform a 2-port OSL calibration at VNA Port 1 and
VNA Port 2.
Vector Network Analyzer MG
PN: 10580-00289 Rev. K
5-10 Phase Measurements
5-19